Helmholtz Coil Pair

Two identical coaxial coils separated by their radius produce a remarkably uniform field between them, the workhorse of magnetometer calibration and Earth-field cancellation. Vary the spacing to see why d = R is special.

Parameters

Coil radiusRRmm
Spacing ratiod/Rd/R1.00 is the Helmholtz condition
Turns per coilNN
CurrentIIA
d = 150 mmR = 150
Fig. 8Helmholtz (d = R). Coils are drawn in perspective; field lines schematic.
Centre field
Centre fieldB0B_0599 µT
In gauss5.995 G
Ideal formula(4/5)3/2μ0NI/R(4/5)^{3/2}\mu_0NI/R599 µT
±1 % region94.1 mm
Spacingdd150 mm
H at centreH0H_0477 A/m
Earth field ratio12 ×
☞ Field within ±1 % of centre over 94.1 mm along the axis (0.627 R).

Axial field of the pair and of each coil

±1 %-200-150-100-500501001502000100200300400500600Position on axis z (mm)B (µT)
paircoil 1coil 2

How it works

Theory · assumptions · formulas

Superposing two loops

Each coil of NN turns at z=±d/2z = \pm d/2 contributes the single-loop axial field, and the pair simply adds:

B(z)=μ0NIR22[1(R2+(z−d2)2)3/2+1(R2+(z+d2)2)3/2]B(z) = \frac{\mu_0 N I R^2}{2}\left[\frac{1}{\bigl(R^2 + (z - \tfrac d2)^2\bigr)^{3/2}} + \frac{1}{\bigl(R^2 + (z + \tfrac d2)^2\bigr)^{3/2}}\right]

Why d = R

By symmetry all odd derivatives of B(z)B(z) vanish at the centre. The second derivative vanishes too when d=Rd = R, leaving the fourth-order term as the leading error, so the field is "maximally flat". At this spacing

B0=(45)3/2μ0NIR≈0.7155 μ0NIRB_0 = \left(\frac{4}{5}\right)^{3/2} \frac{\mu_0 N I}{R} \approx 0.7155\,\frac{\mu_0 N I}{R}

Closer spacing gives a central peak, wider spacing a dip. Try the slider and watch the ±1 % band shrink in both directions.

Practical sizesThe uniform region is roughly a sphere of diameter ≈ 0.3–0.4 R for 1 %. To null the Earth's ≈ 50 µT over a small sensor, a pair of 15 cm radius with 100 turns needs only about 90 mA.